R1500T

High Temperature Transportable blackbody Reference source

LandCAL R1500T Reference Calibration SourceLANDCAL R1500T is a high temperature, transportable blackbody radiation source designed for calibration of a range of infrared thermometers such as System 4 and Fibroptic thermometers. R1500T is a reference standard source giving a resolution of ±1°C.

The LANDCAL R1500T is a high stability, transportable, blackbody, reference source designed for on-site or laboratory calibration of infrared radiation thermometers up to temperatures of 1500 °C.

Six silicon carbide elements heat the conical ended cylindrical blackbody cavity to 1450 °C in approximately 30 minutes. A 3-term controller holds the set temperature to within ±1K.

The R1500T can be used as a transfer standard, providing calibration by the comparison method, using optional standard radiation thermometers. Alternatively a traceable calibration certificate can be supplied for the source, as an optional extra, where direct traceability is required.

Specifications LandCAL R1500T Reference Source

Maximum temperature 1500 °C
Recommended temperature 500 to 1500 °C
Heating rate 30 min. to 1450 °C
Stability <±1K over 30 minutes at set temperature
Radiation cavity Type
Diameter
External aperture
Silicon carbide, 120° cone
45mm dia x 100mm
40mm dia
Emissivity Approx 0.99 at short wavelengths
Heating elements (6) SiC 151/356/20/25.4/7.4
Control thermocouple Pt 13% Rh/Pt, Type R
Controller Eurotherm with RS232C serial interface
Power requirement 110/120V a.c. or 220/240V a.c., 50 to 60Hz
Power consumption 3.0kVA
Overall Dimensions 500 x 380 x 540mm (L x W x H)
Weight 26kg (nett) / 32kg (gross)
Uncertainty (400 to 1500°C) ±3K/6°F (with traceable certificate)

 More Information about the R1500T

 LandCAL Reference Source LandCAL Calibration Sources – Datasheet

Comments are closed.